PROBE DEPLOYMENT
Guidelines for Installation & Sampling Site Maintenance
• Select a water-sampling site that will allow collection of representative water samples.
Position the probe so the sensor surfaces face toward the flow, this will minimize air bubble or fluid
cavitation. Limit flow rate to moderate.
Mount the probe at a 0 to 45° angle. This will prevent the pH (pH/ORP) sensor from becoming
electrically discontinuous due to internal electrolytes flowing away from their internal cells.
• Ensure sampling sites have easy access.
Regularly visit water sampling sites to check for damage to sensors, the installation mountings,
and the meter battery power.
• Remove aquatic weed growth that may be interfering with water sample collection.
Position suspended probes behind a support and anchor the cable/probe to a pipe to protect against
debris.
• Have access to spare sensors and calibration standards or buffers.
• Strictly follow the established standard operating procedures (SOPs).
• Avoid trapped air and maintain constant flow rate if installed in a flow cell.