27
PARAMETER SETUP
Auto: The meter automatically chooses the range (µS/cm or mS/cm) to
optimize the measurement.
Auto mS/cm: The meter automatically chooses the resolution to optimize the
measurement, readings will be in mS/cm only.
Specified numeric resolution: The meter will not autorange, the measurement will be displayed with
the selected resolution only.
Note: A small letter “A” added to the μS/cm or mS/cm unit refers to an absolute conductivity
value.
TDS Resolution
Option: Auto, Auto ppt, 1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt, 1 ppt
Press Modify and use the arrow keys to select the desired TDS resolution.
Press Select to confirm or the ESC key to return to the previous screen.
Auto: The meter automatically chooses the range (ppm or ppt) to optimize
the measurement.
Auto ppt: The meter automatically chooses the resolution to optimize the
measurement, readings will be in ppt only.
Specified numeric resolution: The meter will not autorange, the measurement will be displayed with
the selected resolution only.
7.3. PARAMETER COEFFICIENTS
EC Reference Temperature
Option: 20 °C or 25 °C
This value is used for temperature compensated conductivity. All EC measurements will be referenced
to the conductivity of a sample at this temperature.
Press the functional key to select the desired EC reference temperature.