26
PARAMETER SETUP
Seawater Sigma (σ) Unit
Option: σ
t
, σ
0
, σ
15
Seawater sigma is calculated from the conductivity measurement and depends on water pressure,
temperature, and salinity.
Press the functional key to select the desired reference temperature (current temperature, 0 °C or 15 °C).
EC Resolution
Option: Auto, Auto mS/cm, 1 μS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm, 1 mS/cm
Press Modify and use the arrow keys to select the desired EC resolution.
Press Select to confirm or the ESC key to return to the previous screen.
Auto: The meter automatically chooses the range (µS/cm or mS/cm) to
optimize the measurement.
Auto mS/cm: The meter automatically chooses the resolution to optimize the
measurement, readings will be in mS/cm only.
Specified numeric resolution: The meter will not autorange, the measurement will be displayed with
the selected resolution only.
Absolute EC Resolution
Option: Auto, Auto mS/cm, 1 μS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm, 1 mS/cm
Press Modify and use the arrow keys to select the desired absolute EC resolution.
Press Select to confirm or the ESC key to return to the previous screen.