6 Declaring the significant parameters
Group Parameter Meaning
DIALOG Password Define password (modify), 0000...9999
(Factory preset password: 0000)
PassStat Define password status:
active=password must be input;
inactive=PME can be operated without entering a password
I.DataS to
I.AddFnc
Access to group via keyboard enabled or disabled.
DATA SET
Recall ? You can load either the factory settings or one of the four
parameter sets that have been stored.
Save ? To protect all the instrument set-ups from power failure, they
can be stored in four parameter sets. Whenever you change
from set-up mode to measuring mode you are asked whether
you want the changes to be stored or not. The data will be
permanently saved if you confirm the security prompt with
”Yes” when you exit from set-up mode.
TRANS-
DUCER
ZeromV/V
Zero kN
1)
Nom.mV/V
NVal kN
1)
Physic. unit
mV/V
0
Zero mV/V
Zero
kN
Nom.mV/V ( 2 mV/V)
Nom.kN
( 10 kN for 2 mV/V)
Transducer characteristics: Nominal value 10 kN;
Rated sensitivity 2 mV/V
Setting up in accordance with transducer characteristics
1
) Depending on the unit chosen