60
PME-MP55
HBM A0563-5.6 en
9 Specifications
Type MP55
Accuracy class 0.1
2)
Supply voltage
Permitted supply voltage range
Power consumption
V
DC
V
DC
W
24; Potential separation from measuring
system, test voltage 350 V
eff
3)
)
18...30
9 maximum
Amplifier
Carrier frequency kHz 4.8"1 %
Bridge excitation voltage U
B
("5 %) V
rms
5 / 2.5 / 1
Transducers that can be connected
SG half and full bridge
Inductive half and full bridge, LVDTs
Ω
mH
220...5000 / 110...5000 / 60...5000
8...160 / 4...160 / 2...160
Permitted cable length between trans-
ducer and amplifier
m max. 500
Maximum permitted common mode
voltage
V ±5
Common mode rejection
0...500 Hz
0...4800 Hz
dB
dB
120
72
Maximum differential voltage mV "30
Linearity deviation (typical) % 0.025
Noise voltage
1)
Measuring range [mV/V]
3 50 500
0...10 Hz
0...500 Hz
µV/V
PP
µV/V
PP
0.2 3 30
1.5 25 250
Measurement frequency range, adjust-
able (-1 dB)
Hz 0.05...500
Max. display resolution 999 999 digits at 6.67 % of amplifier input
range
Min. display resolution 10 digits at 100 % of amplifier input range
Input sensitivities (measuring ranges
can be adjustet by DIP switch)
Low Medium High
at U
B
=5 V
at U
B
=2.5 V
at U
B
=1 V
mV/V
mV/V
mV/V
0.15...3 2.5...50 25...500
0.3...6 5...100 50...1000
0.75...15 12.5...250 125...2500
Low pass filter Adjustable in steps of 0.05 to 500 Hz
(Bessel and Butterworth filter characteris-
tics)
Effect of operating voltage when speci-
fied range is changed (in relation to final
value)
on zero point
on sensitivity
%
%
< 0.01 f.s
< 0.01 f.s
1)
When U
B
=5 V, by reference to the input
2)
0.25 with irradiation per EN61326 in the range of 700 MHz up to 1 GHz
3)
Type-tested per EN61010-1:2001