16
CMD600
A28201.1 en
charge_lost int =0 charge information preserved, range
capacitor was not switched
=1 charge information lost
range int =1 only small capacitor
=2 both capacitors
inquiry CH_GAIN = ?
response OK, CH_GAIN = actual_gain, charge_lost, range
help CH_GAIN ?
response OK, CH_GAIN gain (min = min_gain, max = max_gain)
parameters:
min_gain float minimum gain in V/C
max_gain float maximum gain in V/C
CH_SYS_GAIN
Used to send system gain of amplifier (for preselected channel if applicable). System gain
of amplifier is dependent from ch_gain and sensor sensitivity. This gain is not stored in
EEPROM. This gain actually product between sensor senditivity and channel_gain, and
those two values are stored in EEPROM.
command CH_SYS_GAIN sys_gain
parameters:
sys_gain float requested gain in V/En_UNIT
response OK, CH_SYS_GAIN = actual_sys_gain
parameters:
actual_sys_gain float actual set gain in V/En_UNIT
charge_lost int =0 charge information preserved, range
capacitor was not switched
=1 charge information lost
range int =1 only small capacitor
=2 both capacitors
inquiry CH_SYS_GAIN = ?
response OK, CH_SYS_GAIN = actual_sys_gain, charge_lost, range
help CH_SYS_GAIN ?
response OK, CH_SYS_GAIN sys_gain (min = min_gain, max = max_gain)
parameters:
min_gain float minimum gain in V/En_UNIT
max_gain float maximum gain in V/En_UNIT