18
CMD600
A28201.1 en
CH_OUT_OFFSET
Used to set output offset voltage for preselected channel. Example if output offset
voltage is 0,5 V then charge amplifier will have 0,5 V on output if it is in reset state.
command CH_OUT_OFFSET voltage
parameters:
voltage float Offset voltage in V.
response OK, CH_OUT_OFFSET = voltage
parameters:
voltage float Offset voltage in V
inquiry CH_OUT_OFFSET = ?
response OK, CH_OUT_OFFSET = voltage
help CH_OUT_OFFSET?
response OK, CH_OUT_OFFSET voltage (min = min_voltage, max = max_voltage)
parameters:
min_voltage float Minimum offset voltage in V. This is usually
negative value around 1 V.
max_voltage float Maximum offset voltage in V. This is usually
positive value around 1 V.
CH_VALUE
Used to inquiry preselected channel value
command not applicable
inquiry CH_VALUE= ?
response OK, CH_VALUE = output_voltage, output_charge, overload
parameters:
output_voltage float Output value of charge amplifier in V
output_in_EnUnit float Output value of charge amplifier in En_Unit
overload int 0.. channels was not overloaded from last
reset. Default after reset of channel.
1.. channel was overloaded and charge
information is lost.
help CH_ CH_VALUE?
response OK, CH_VALUE output_voltage, output_in_EnUnit, overload