EasyManua.ls Logo

HBM PACEline CMD600 - Easyteach

HBM PACEline CMD600
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
CMD600
A28201.1 en
response OK, CH_ AUTOSCALE = step
parameters:
step int 0auto scale procedure is aborted or auto
scale procedure is completed 1start auto
scale(channel gain is set to 600nC) 2...
channel gain is set to 60nC. 3...channel gain is
set to 6nC. 4....channel gain is set to 500pC.
5....end of auto scale procedu
help CH_ AUTOSCALE?
response OK, CH_AUTOSCALEstep
EASYTEACH_ALLOW (EasyTeach = SensorTeach)
Enable/disable easyteach function
command EASYTEACH_ALLOW enable
parameters:
enable int 0.. Disable easyteach function with digital input
1.. Enable easyteach function with digital input
response OK, EASYTEACH_ALLOW = enable
parameters:
enable int See description above
inquiry EASYTEACH_ALLOW = ?
response EASYTEACH_ALLOW = enable
help EASYTEACH_ALLOW?
response OK, EASYTEACH_ALLOW enable (0..dissable, 1.. enable)
CH_EASYTEACH (EasyTeach = SensorTeach)
Start easyteach function for preselected channel
command CH_EASYTEACH
response OK, CH_EASYTEACH
inquiry Not aplicable
help CH_EASYTEACH?
response OK, CH_EASYTEACH start easyteach

Related product manuals