17
CMD600
A28201.1 en
CH_SENSOR_SENSITIVITY
Set sensor sensitivity of sensor attached to charge amplifier.
command CH_SENSOR_SENSITIVITY sensitivity
parameters:
sensitivity float requested sensitivity in C/En_Unit
response OK, CH_SENSOR_SENSITIVITY = sensitivity
parameters:
sensitivity float Look description above.
inquiry CH_ SENSOR_SENSITIVITY = ?
response OK, CH_ SENSOR_SENSITIVITY = sensitivity
help CH__ SENSOR_SENSITIVITY ?
response OK, CH__ SENSOR_SENSITIVITY sensitivity(C/ En_Unit)
CH_HPF
Used to configure high pass filter on preselected channel
command CH_HPF corner_ freq
parameters:
corner_ freq float corner frequency in Hz of high pass filter
(3dB). 0 means that high pass filter is disabled
response OK, CH_HPF = corner_ freq
parameters:
corner_ freq float actual corner frequency in Hz
inquiry CH_HPF = ?
response OK, CH_HPF = corner_ freq
help CH_HPF?
response OK, CH_HPF corner_ freq (freq1, freq2, )
parameters:
freq1.freq_n float list of all possible high pass filter frequencies
CH_LPF
Used to configure low pass filter on preselected channel.
command CH_LPF corner_ freq
parameters:
corner_ freq float corner frequency in Hz of low pass filter (3dB).
response OK, CH_LPF = corner_ freq
parameters:
corner_ freq float actual corner frequency in Hz
inquiry CH_LPF = ?
response OK, CH_LPF = corner_ freq
help CH_LPF?
response OK, CH_LPF corner_ freq (freq1, freq2, )
parameters:
freq1.freq_n float list of all possible low pass filter frequencies