EasyManua.ls Logo

HBM PACEline CMD600 - Input_Offset; Cal_Dac; Ad_Data

HBM PACEline CMD600
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38
CMD600
A28201.1 en
CH_INPUT_OFFSET
Used to set input offset for first amplifier (integrator). Value is in DAC quants and
corresponds to +/3mV. This parameter isn’t changed when factory default is carried out.
command CH_INPUT_OFFSET select_input_offset,input_offset
parameters:
select_input
_ offset
int 0...input offset, when high pass filter is off.
1...input offset for slow high pass filter. 2...input
offset for fast high pass filter..
input_offset int Input offset in quants.
response OK, CH_INPUT_OFFSET = select_input_offset,input_offset
parameters:
select_input
_ offset
int Look decsription above.
input_offset int Input offset in quants.
inquiry CH_ INPUT_OFFSET = ?
response OK, CH_ INPUT_OFFSET = input_offset(HPF off), input_offset(slow HPF),
input_offset(fast HPF)
help CH_ INPUT_OFFSET ?
response OK, CH_ INPUT_OFFSET select_input_offset(0..2),input_offset
(min = 2048, max = 2047)
CH_CAL_DAC
Used for DAC calibration. Sets output dac value in quants.
command CH_CAL_DAC dac_quants
parameters:
dac_quants int DAC quants(04095)
response OK, CH_ CAL_DAC = dac_quants
parameters:
dac_quants int See description above.
help CH_ CAL_DAC ?
response OK, CH_ CAL_DAC dac_quants (min = 0, max = 4095)
AD_DATA
Return average last 50 AD
streaming is enabled.
value in quants. This command can not be use when data
command Not applicable
command AD_DATA=?
response OK, AD_DATA = ad_quants
parameters:
ad_quants int Average last 50 AD converter value in quants.
help AD_DATA?
response OK, AD_DATA