9. Quality control testing
Test at least one known negative and one known positive specimen or control,
whenever a new bottle of strips is rst opened, for each new shipment, for each
new LOT or at least monthly. Quality control samples should be run like a patient
sample. Do not use water as negative control. Positive and negative control
solutions provide a convenient basis for a quality control program. Contact the
service number for ordering information.
If proper results are not obtained, consult your local product representative or
contact Customer Service by calling the number given at the end of this manual
for advise on testing techniques and results.
9.1. How to run a quality control test
Prepare the check urines as described in the package insert and test them in
test mode.
Handle the check solutions exactly as patient samples.
Upon pressing
?
on the start menu the "CHECK MODE" screen will appear.
MENU
Memory
Settings
Check mode
Display 18: Main menu
CHECK MODE
Memory
Measurement
Display 19: Check mode
By pressing the panel the sample selection will appear. Select the type of
control sample you want to analyze.