11. Cleaning and Maintenance
DANGER:
Urine and used test strips bare the danger of infection. Always use protective
gloves during handling and disposal. The disposal of used test strips should
be performed according to the regulations for the handling of potentially infec-
tious material
11.1. How to clean the housing
The instrument housing may be wiped with a cloth. Mild cleaning agents or
disinfectants may be used. Ensure that no moisture permeates the equipment
11.2. How to clean the strip holder
Wipe off urine residues from the strip holder with a lint-free cloth after each mea-
surement. This prevents crustication and drying of urine residues.
The strip holder can be removed from its transport mechanism and should be
cleaned with water and - when necessary - with cleaning agent or disinfectant.
Make sure that the instrument is turned off before removing the test slide.
After cleaning, the strip holder should be put back onto its transport mechanism
carefully. The rectangular notches of transport mechanism and strip retainer
must be placed on top of each other (Pic. 14).
Pic. 12: Test slide
(bottom view)
Pic. 13: Notch A Pic. 14: Notch B