5. Precautions, Limitations and Warnings
The Element COAG+ Analyzer is for in vitro diagnostic use
only. Some environmental factors and incorrect operation
may cause the Element COAG+ Analyzer to not work
effectively or to report inaccurate results. Please read the
following warnings carefully before use and strictly follow
them.
As with all diagnostic tests, Element COAG+ Analyzer’s test
results should be analyzed and interpreted by combining the
condition, medical history, current and possible treatment of
a specific animal. Any results exhibiting inconsistency with
the clinical status should be repeated or supplemented with
additional test data or repeated with other testing methods.
5.1 Element COAG+™ Analyzer Care
●
DO NOT spill any liquid on the Element COAG+ Analyzer. In this
case, immediately contact Heska’s Technical Support Services.
●
The Element COAG+ Analyzer is a delicate instrument and
should be handled with care. Dropping or mishandling may cause
malfunction of Element COAG+ Analyzer.
●
The Element COAG+ Analyzer should be transported in a carrying
case or a secure container.
●
DO NOT store the Element COAG+ Analyzer below 14°F or above
104°F (-10ºC or above 40ºC).
●
DO NOT use the Element COAG+ Analyzer with any other test
strips not provided by Heska.
5.2 Test Environment Requirements
●
The temperature should be maintained between 50°F-95°F (10ºC
and 35ºC).
●
This equipment has been tested and found to comply with
applicable EMC emission requirements as specified in EN 55032.
The emissions of the energy used are low and not likely to cause
interference in nearby electronic equipment.
●
Do not use this instrument in a dry environment, especially if
synthetic materials are present. Synthetic clothes, carpets, etc.,
may cause static discharges in a dry environment. That may cause
incorrect operation or damage to the device. The recommended
humidity operating range is 10% to 90%.