Specifications
Electrical
Description
Chapter
1:
Introduction
For
Isolette®
Infant
Incubator
electrical
specifications,
see
table
1-2
on
page
1-
11.
Table
1-2.
Electrical
Specifications
Feature
Dimension
Power
requirements
(100V
fixed
pedestal
stand
model)
100V
+
10%,
50/60
Hz,
990
W
maximum
Power
requirements
(120V
fixed
pedestal
stand
model)
120V
+
10%,
60
Hz,
1188
W
maximum
Power
requirements
(220V
and
240V
fixed
pedestal
stand
model)
220V
or
240V
+
10%,
50/60
Hz,
2376
W
maximum
Power
requirements
(100V
variable
height
adjustable
pedestal
stand
model)
100V
+
10%,
50/60
Hz,
990
W
maximum
Power
requirements
(120V
variable
height
adjustable
pedestal
stand
model)
120V
+
10%,
60
Hz,
1188
W
maximum
Power
requirements
(220V
and
240V
variable
height
adjustable
pedestal
stand
model)
220V
or
240V
+
10%,
50/60
Hz,
2376
W
maximum
Convenience
outlets
(100V
model)
100V,
100
W
maximum
Convenience
outlets
(120V
model)
120V,
300
W
maximum
Convenience
outlets
(220V
and
240V
220V
or
240V,
500
W
model)
models)
maximum
Chassis
current
leakage
(110V
and
120V
|
300
uA
or
less
model)
Chassis
current
leakage
(220V
and
240V
|
500
uA
or
less
Isolette®
Infant
Incubator
Service
Manual
(man223rf)
Page
1-11