Loading...
Specifications
Chapter
1:
Introduction
Oxygen
Control
Description
For
Isolette®
Infant
Incubator
oxygen
control
specifications,
see
table
1-3
on
page
1-12.
Table
1-3.
Oxygen
Control
Specifications
Feature
Dimension
Oxygen
display
range
18%
to
100%
Oxygen
display
resolution
1%
oxygen
Oxygen
display
accuracy
(100%
+
3%
calibration)
Oxygen
display
accuracy
(21%
calibration);
+
5%
Oxygen
set
resolution
1%
Regulations,
Standards,
and
Codes
With
respect
to
the
International
Electrotechnical
Commission
(IEC)
601-1,
the
incubator
is
Class
1,
Type
BF.
The
mode
of
operation
is
continuous
operation
with
short-time
loading
(variable
height
adjustable
model
only).
Page
1-12
Isolette®
Infant
Incubator
Service
Manual
(man223rf)

Table of Contents

Related product manuals