Final
Actions
Chapter
2:
Troubleshooting
Procedures
Yes
No
y
->
The
oxygen
concentration
level
is
too
high.
Go
to
RAP
2,5.
32.
Check
the
humidity
system
accessory
(if
applicable).
a.
Fill
the
reservoir,
and
place
the
probe
of
a
calibrated
hygrometer
inside
the
hood
(A)
at
the
center
mattress
(L).
b.
Pre-warm
the
incubator
to
95°F
(35°C).
On
Display
1,
press
the
Humidity
softkey.
d.
Turn
on
the
humidity.
Set
the
humidity
setpoint
to
50%.
e.
Within
30
min
to
60
min,
the
hygrometer
and
Humidity
display
reads
50%
+
6%.
Yes
No
|
y
>
GotoRAP
2.1.
33.
Go
to
“Final
Actions”
on
page
2-14.
| |
Final
Actions
1.
Complete
the
required
preventive
maintenance
procedures.
See
"Preventive
Maintenance
Checklist”
on
page
6-8.
2.
Complete
all
required
administrative
tasks.
Page
2-14
Isolette®
infant
Incubator
Service
Manual
(man223r1)