Hill-Rom Isolette C2000 - Oxygen Concentration is Low

Hill-Rom Isolette C2000
93 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.4
Oxygen
Concentration
is
Low
Chapter
2:
Troubleshooting
Procedures
2.4
Oxygen
Concentration
is
Low
1.
The
oxygen
source
is
connected
to
the
incubator.
Yes
No
J
Connect
the
oxygen
to
the
incubator.
If
this
solves
the
problem,
go
to
“Final
Actions”
on
page
2-14.
Otherwise,
go
to
step
2.
2.
Check
all
access
doors
(A)
and
iris
entry
ports
(B)
(see
figure
2-14
on
page
2-38).
Figure
2-14.
Access
Doors
and
Iris
Entry
Ports
m2231043
All
access
doors
(A)
and
iris
entry
ports
(B)
are
closed.
Yes
No
J
Close
all
access doors
(A)
and
iris
entry
ports
(B). If
this
solves
the
problem,
go
to
“Final
Actions”
on
page
2-14.
Otherwise,
go
to
step
3.
3.
Check
all
iris
entry
port
(B)
sleeves.
All
iris
entry
port
(B)
sleeves
are
properly
installed.
Page
2
-38
lsolette®
Infant
Incubator
Service
Manual
(man223rf)