2.4
Oxygen
Concentration
is
Low
Chapter
2:
Troubleshooting
Procedures
2.4
Oxygen
Concentration
is
Low
1.
The
oxygen
source
is
connected
to
the
incubator.
Yes
No
J
Connect
the
oxygen
to
the
incubator.
If
this
solves
the
problem,
go
to
“Final
Actions”
on
page
2-14.
Otherwise,
go
to
step
2.
2.
Check
all
access
doors
(A)
and
iris
entry
ports
(B)
(see
figure
2-14
on
page
2-38).
Figure
2-14.
Access
Doors
and
Iris
Entry
Ports
m2231043
All
access
doors
(A)
and
iris
entry
ports
(B)
are
closed.
Yes
No
J
Close
all
access doors
(A)
and
iris
entry
ports
(B). If
this
solves
the
problem,
go
to
“Final
Actions”
on
page
2-14.
Otherwise,
go
to
step
3.
3.
Check
all
iris
entry
port
(B)
sleeves.
All
iris
entry
port
(B)
sleeves
are
properly
installed.
Page
2
-38
lsolette®
Infant
Incubator
Service
Manual
(man223rf)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hill-Rom Isolette C2000 and is the answer not in the manual?

Hill-Rom Isolette C2000 Specifications

General IconGeneral
BrandHill-Rom
ModelIsolette C2000
CategoryMedical Equipment
LanguageEnglish

Related product manuals