System
Failure
Messages
Chapter
2:
Troubleshooting
Procedures
2.5
Oxygen
Concentration
is
High
1.
Check
the
air
intake
microfilter
(A)
(see
figure
2-16
on
page
2-41).
Figure
2-16.
Air
Intake
Microfilter
and
Fan
Impeller
m223f045
The
air
intake
microfilter
(A)
is
clean.
Yes
No
L
>
Replace
the
air
intake
microfilter
(A)
(refer
to
procedure
4.4).
If
this
solves
the
problem,
go
to
“Final
Actions”
on
page
2-14.
Otherwise,
go
to
step
2.
2.
Check
the
fan
impeller
(B).
The
fan
impeller
(B)
is
clean.
Yes
No
|
Clean
the fan
impeller
(B).
If
necessary,
replace
the
fan
impeller
(B).
If
this
solves
the
problem,
go
to
“Final
Actions”
on
page
2-14.
Otherwise,
call
Hill-Rom
Technical
Support
at
(800)
445-3720
for
assistance.
3.
Call
Hill-Rom
Technical
Support
at
(800)
445-3720
for
assistance.