Safety
Tips
Chapter
1:
Introduction
A
A
CAUTION:
When
reconnecting
the
rear
panel
connectors,
make
sure
the
sensor
module
cable
connector
is
connected
to
the
sensor
module
connector,
and
not
the
RS-232
connector.
CAUTION:
|
Replace
both
oxygen
sensor
cells
at
the
same
time.
Failure
to
do
so
could
result
in
equipment
damage.
CAUTION:
Do
not
use
harsh
cleaners,
solvents,
or
detergents.
Equipment
damage
could
occur.
CAUTION:
Some
chemical
cleaning
agents
may
be
conductive
and/or
leave
a
residue
that
may
permit
a
build-up
of
dust
or
dirt,
which
may
be
conductive.
Do
not
permit
cleaning
agents
to
contact
electrical
components.
Do
not
spray
cleaning
solutions
onto
any
of
these
surfaces.
CAUTION:
When
cleaning
the
interior
of
the
incubator
shell,
care
must
be
taken
to
prevent
liquids
from
entering
the
motor
shaft
opening.
Equipment
damage
could
occur.
CAUTION:
Alcohol
can
cause
crazing
of
the
clear
acrylic
hood.
Do
not
use
alcohol
for
cleaning.
CAUTION:
Do
not
expose
the
hood
assembly
to
direct
radiation
from
germicidal
lamps.
Ultraviolet
radiation
from
these
sources
can
cause
cracking
of
gaskets,
fading
of
paint,
and
crazing
of
the
clear
acrylic
hood.
Page
1-18
IsoletteĀ®
Infant
Incubator
Service
Manual
(man223rf)