5.3 Measurement specifications
140
Measurement accuracy
Values for f.s. depend on measurement ranges.
“F” in the tables refers to the frequency in kHz.
Add ±1 mA to DC measurement accuracy for current.
Add (±1 mA) × (voltage read value) to DC measurement accuracy for active power.
When using the 200 mA or 500 mA range, add ±0.1% rdg. to current and active
power for which 1 kHz < f 10 kHz.
Values for voltage, current, and active power for which 0.1 Hz f < 10 Hz are for ref-
erence only.
Values for voltage, current, and active power in excess of 220 V or 20 A for which 10
Hz f < 16 Hz are for reference only.
Values for current and active power in excess of 20 A for which 500 Hz < f 50 kHz
are for reference only.
Values for current and active power in excess of 15 A for which 50 kHz < f 100 kHz
are for reference only.
Values for voltage and active power in excess of 750 V for which 30 kHz < f 100
kHz are for reference only.
Effective measuring range Voltage: 1% to 130% of range (1,000 V range, up to 1,000 V)
Current: 1% to 130% of range
Active power: 0% to 169% of range (when using 1,000 V range, up to 130%)
However, defined when voltage and current fall within the effective measurement range.
Other parameters: Valid within the voltage, current, and active power effective mea-
surement range.
Voltage
Frequency(f) Input < 50% f.s. 50% f.s. Input < 100% f.s. 100% f.s. Input
DC
±0.1%rdg.±0.1%f.s. ±0.1%rdg.±0.1%f.s. ±0.2%rdg.
0.1 Hz f < 16 Hz ±0.1%rdg.±0.2%f.s. ±0.3%rdg. ±0.3%rdg.
16 Hz f < 45 Hz ±0.1%rdg.±0.1%f.s. ±0.2%rdg. ±0.2%rdg.
45 Hz f 66 Hz ±0.1%rdg.±0.05%f.s. ±0.15%rdg. ±0.15%rdg.
66 Hz < f 500 Hz ±0.1%rdg.±0.1%f.s. ±0.2%rdg. ±0.2%rdg.
500 Hz < f 10 kHz ±0.1%rdg.±0.2%f.s. ±0.3%rdg. ±0.3%rdg.
10 kHz < f 50 kHz ±0.5%rdg.±0.3%f.s. ±0.8%rdg. ±0.8%rdg.
50 kHz < f 100 kHz ±2.1%rdg.±0.3%f.s. ±2.4%rdg. ±2.4%rdg.
Current (direct input)
Frequency(f) Input < 50% f.s. 50% f.s. Input < 100% f.s. 100% f.s. Input
DC ±0.1%rdg.±0.1%f.s. ±0.1%rdg.±0.1%f.s. ±0.2%rdg.
0.1 Hz f < 16 Hz ±0.1%rdg.±0.2%f.s. ±0.3%rdg. ±0.3%rdg.
16 Hz f < 45 Hz ±0.1%rdg.±0.1%f.s. ±0.2%rdg. ±0.2%rdg.
45 Hz f 66 Hz ±0.1%rdg.±0.05%f.s. ±0.15%rdg. ±0.15%rdg.
66 Hz < f 500 Hz ±0.1%rdg.±0.1%f.s. ±0.2%rdg. ±0.2%rdg.
500 Hz < f 1 kHz ±0.1%rdg.±0.2%f.s. ±0.3%rdg. ±0.3%rdg.
1 kHz < f 10 kHz ±(0.03+0.07×F)%rdg.±0.2%f.s. ±(0.23+0.07×F)%rdg. ±(0.23+0.07×F)%rdg.
10 kHz < f 100 kHz
±(0.3+0.04×F)%rdg.±0.3%f.s.
±(0.6+0.04×F)%rdg.
±(0.6+0.04×F)%rdg.
Active power
Frequency(f) Input < 50% f.s. 50% f.s. Input < 100% f.s. 100% f.s. Input
DC ±0.1%rdg.±0.1%f.s. ±0.1%rdg.±0.1%f.s. ±0.2%rdg.
0.1 Hz f < 16 Hz ±0.1%rdg.±0.2%f.s. ±0.3%rdg. ±0.3%rdg.
16 Hz f < 45 Hz ±0.1%rdg.±0.1%f.s. ±0.2%rdg. ±0.2%rdg.
45 Hz f 66 Hz ±0.1%rdg.±0.05%f.s. ±0.15%rdg. ±0.15%rdg.
66 Hz < f 500 Hz ±0.1%rdg.±0.1%f.s. ±0.2%rdg. ±0.2%rdg.
500 Hz < f 1 kHz ±0.1%rdg.±0.2%f.s. ±0.3%rdg. ±0.3%rdg.
1 kHz < f 10 kHz ±(0.03+0.07×F)%rdg.±0.2%f.s. ±(0.23+0.07×F)%rdg. ±(0.23+0.07×F)%rdg.
10 kHz < f 50 kHz
±(0.07×F)%rdg.±0.3%f.s.
±(0.3+0.07×F)%rdg.
±(0.3+0.07×F)%rdg.
50 kHz < f 100 kHz
±(0.6+0.07×F)%rdg.±0.3%f.s.
±(0.9+0.07×F)%rdg.
±(0.9+0.07×F)%rdg.