3.4 Viewing Harmonic Measured Values
72
The following two harmonic display methods are available:
Displaying components for the same order as the display parameter (default state)
Allocating components for different orders to areas a, b, c, and d on the display
Example: If [odr 1] or similar is shown in area a on the display while displaying harmonics
1 Press to display LEVEL.
2 Press or to change the
order.
The component for the order that is
specified in the area a will be shown for
areas b, c, and d on the display.
Settings: 0 to 50
1 Press after pressing to
place the instrument in the shift state.
2 Press or to change the
order.
If all display areas indicate [--], the “display the
same order component” setting will be used.
Settings: 0 to 50 → - - → 0 to 50
To move areas a, b, c, or d on the dis-
play :
Press or to select the display area
you wish to set so that it flashes and then set
as desired.
3 Press or to exit the Order
Setting screen.
By allocating different orders to areas a
through d on the display and then setting all
the areas to the same measurement parame-
ter, you can observe changes in each order.