3.4 Viewing Harmonic Measured Values
73
Total harmonic voltage distortion (THD V %), total harmonic current distortion (THD A %)
Fundamental wave voltage (FND V), fundamental wave current (FND A), fundamental wave active power
(FND W), fundamental wave apparent power (FND VA), fundamental wave reactive power (FND var), funda-
mental wave power factor (FND PF), fundamental voltage/current phase difference (FND °), inter-channel
voltage fundamental wave phase difference (FND θ V °), inter-channel current fundamental wave phase dif-
ference (FND θ A °)
Parameters shown with normal display parameters
Example: THD V%
Press through to display THD V
% or THD A % in the display area.
Parameters shown as fundamental wave parameters (displayed as RECTIFIER FND)
Measurement parameters displayed using harmonic analysis (harmonic level, content percentage, rectifier
FND [fundamental wave component]) are not averaged by the averaging function.
1 Press to place the instrument
in the shift state and then press
to display FND.
2 Press through to change
the display parameter.
Example: FND V