Note: If the hose is not blocked and unblocked within 2.5 minutes, the detector
will assume the pump has failed and deactivates.
Successful Pump Test: If the pump test is successful, the following screen
displays.
Unsuccessful Block Pump Test: If the pump is not operating correctly, the
following screens display before the detector deactivates.
Activate the detector again. If the pump fails the startup again, refer to See
"Troubleshooting" on page98 for more information.
If the diffusion cover is not attached, the detector beeps and the following
screen displays before continuing with the startup tests.
When the diffusion cover is replaced, the detector activates the pump
alarm.
BW MaxXT II 18 User Manual