Plum A+
®
Infusion Systems
430-95597-001 (Rev. 06/05) 4- 9
2 Enter the desired Rate using
the keypad.
NOTE: When Rate is changed, the
duration is automatically recomputed
using the current VTBI. To
understand the Automatic Calculation
feature, see Examples of Automatic
Calculation on page 6-10.
3 Press .
CAUTION: VERIFY ALL VALUES
BEFORE STARTING INFUSION.
Program Confirmation screen
will appear. If correct, press
[Yes] to begin infusion.
NOTE: During Titration, infuser
continues to deliver at previous parameters until new program is
confirmed.
Example of Main Delivery
Screen after being titrated.
NOTE: Volume Infused is an
incrementing value which is cleared
in
[Options/Vol Inf].
The following 3 screens are
examples of changed values.
A
PROGRAM
Rate
VTBI
Duration
mL/hr
mL
hr : min
Program
Options
Cancel/
Back
Drug
List
Enter Value using keypad
15
50
03:20
MedSurg
A
PROGRAM
Drug
Rate
VTBI
Delivery will be:
15
50
mL/hr
mL
Confirm Program?
No Drug Selected
!
Yes No
MedSurg
A
A
B
B
PUMPING PUMPING
Rate
mL/hr
Vol Inf
mL
Options/
Vol Inf
15
41.
6
20
50
!
!
MedSurg
sec04.fm Page 9 Wednesday, June 8, 2005 8:22 AM