Plum A+
®
Infusion Systems
430-95597-001 (Rev. 06/05) 4- 17
understand the Automatic Calculation feature, see Examples of
Automatic Calculation on page 6-10.
6 Use to move down to the DURATION field.
(Optional) Enter values for desired hours and
minutes.
NOTE: When Duration is entered, Rate is automatically
computed.
7 Press .
CAUTION: VERIFY ALL VALUES BEFORE STARTING INFUSION.
Program Confirmation screen will appear. If correct,
press
[Yes] to begin infusion.
Example of a concurrent
delivery.
A
A
B
B
PUMPING PUMPING
Rate
mL/hr
Vol Inf
mL
Options/
Vol Inf
5.
5
20.
4
1
10
! !
MedSurg
sec04.fm Page 17 Wednesday, June 8, 2005 8:22 AM