EasyManua.ls Logo

HP 54645A - To Autoscale on a Video Signal

HP 54645A
301 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
To autoscale on a video signal
1
Use a cable to connect a TV signal to channel 1.
2
Press
Slope/Glitch
in the
TRIGGER
section of the front panel,
and select the
TV
softkey as your trigger mode.
3
To select a TV standard, press the
TV Menu
softkey, then press the
Standard
softkey to select the TV standard. Your choices are
NTSC
,
PAL
,
SECAM
, and
GENERIC
.
GENERIC
is used for other TV/Video
standards. If your TV standard has been previously selected, you
may skip this step.
4
Press
Slope/Glitch
, then press the
Video Autoscale
softkey.
Provide Correct Matching
Many TV signals are produced from 75- sources. To provide correct matching
to these sources, an HP 11094B 75- load is included as an accessory. For
oscilloscopes that have selectable input impedance, the 1-M input should be
used with the 75- load.
The
Undo Autoscale
softkey in the Setup menu resets the instrument to the
configuration that was present before pressing
Video Autoscale
.
Using Option 005 Enhanced TV/Video Trigger (HP 54645A)