HP 54645A - To Verify Bandwidth

HP 54645A
301 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
To verify bandwidth
In this test you verify bandwidth by using a power meter and power sensor to
set output of a signal generator at 1 MHz and at 100 MHz. You use the
peak-to-peak voltage at 1 MHz and at 100 MHz to verify the bandwidth
response of the oscilloscope. The test is repeated at 5 mV/div (75-MHz
bandwidth).
Test limits at 10 mV/div to 5 V/div:
all channels (
±
3 dB)
dc to 100 MHz
ac coupled 10 Hz to 100 MHz
Test limits at <10 mV/div:
all channels (
±
3 dB)
dc to 75 MHz
ac coupled 10 Hz to 75 MHz
Table 7-7
Equipment Required
Equipment Critical specifications Recommended Model/Part
Signal generator 1 to 100 MHz at 200 mV HP 8656B opt 001
Power meter and
Power Sensor
1 to 100 MHz
±
3% accuracy HP 436A and HP 8482A
Power splitter Outputs differ by < 0.15 dB HP 11667B
Cable Type N (m), 24 inch HP 11500B
Adapter Type N (m) to BNC (m) HP 1251-0082
Feedthrough 50
, BNC (m) and (f) HP 11048C
Testing, Adjusting, and Troubleshooting
To verify bandwidth
7-19