EasyManua.ls Logo

HP 54645A - To Verify Horizontal T and 1; T Accuracy

HP 54645A
301 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
To verify horizontal t and 1/t accuracy
In this test, you verify the horizontal
t and 1/
t accuracy by measuring the
output of a time mark generator with the oscilloscope.
Test limits:
±
0.01%
±
0.2% of full scale
±
40 ps (same channel)
Table 7-8
Equipment Required
Equipment Critical specifications Recommended Model/Part
Time marker generator Stability 5 ppm after 1/2 hour TG 501A and TM 503B
Cable BNC, 3 feet HP 10503A
Termination 50
, BNC connectors (m) and (f) HP 11048C
1
Connect the time mark generator to channel 1 using the 50-
feedthrough at the oscilloscope input. Then, set the time mark
generator for 0.1-ms markers.
Channel 1 is labeled "1" on HP 54645A and "A1" on HP 54645D.
2
Set up the oscilloscope.
a
Press
Setup
, then press the
Default Setup
softkey.
a
Press
Display
, then press the
Vectors Off
softkey.
c
Press
Autoscale
.
d
Set the time base to 20
µ
s/div.
e
Press
Main/Delayed
, then press the
Time Ref Left
softkey.
f
Adjust the trigger level to obtain a stable display.
3
Press
Time
, then press the
Freq
and
Period
softkeys.
You should measure the following:
Frequency 10 kHz, test limits are 9.96 kHz to 10.04 kHz.
Period 100
µ
s, test limits are 99.59
µ
s to 100.41
µ
s.
If the measurements are not within the test limits, see "Troubleshooting the
Oscilloscope."
4
Change the time mark generator to 1
µ
s, and change the time base to
200 ns/div. Adjust the trigger level to obtain a stable display.
Testing, Adjusting, and Troubleshooting
To verify horizontal
t and 1/
t accuracy
7-22

Table of Contents