HP 54645A - HP 54645 A;D Oscilloscopes

HP 54645A
301 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HP 54645A/D Oscilloscopes
Number of Channels 2 Analoglabeled 1 and 2 on HP 54645A,
A1 and A2 on HP 54645D
Probes 10074A
Range 1 mV/div - 5 V/div
Vernier calibrated ± 3%
Position (offset) range ± 8 divisions minimum
± 2V on ranges < 200 mV/div
± 40V on ranges > 200 mV/div
Dynamic Input Range Lesser of ± 8 div or ± 32 V
DC Vertical Gain Accuracy
1, 2, 5 sequences
Vernier
± 1.5% of full scale
± 3% of full scale
DC Vertical Offset Accuracy
serial prefix < US3707
serial prefix US3707
± 1% of full scale ± 0.5% of position value
± 1% of full scale ± 0.5% of position value (10 mV/div)
± 2% of full scale ± 0.5% of position value (< 10 mV/div)
Single-Cursor Accuracy DC vertical gain accuracy + DC vertical offset
accuracy ± 1/2 LSB (LSB=0.4% of full scale)
Dual-Cursor Accuracy DC vertical gain accuracy ± 1 LSB
Bandwidth (3dB)
Repetitive
Single shot
100 MHz @ 10 mV/div (75 MHz @ < 10 mV/div)
50 MHz
Bandwidth limit (~ 20 MHz)
Rise Time (calculated) ~3.5 ns @ > 10 mV/div
~3.9 - 4.6 ns @ < 10 mV/div
Coupling AC, DC, GND
AC Coupling Corner Frequency ~1.5 Hz
Input Impedance 1 M, ± 1%, ~13 pF
Maximum Input 400 V (DC + peak AC)
Probe ID (HP & Tek compatible) 1X, 10X, 20X, 100X
ESD Tolerance ± 2 kV
Channel Isolation (with channels at
the same v/div)
DC to 20 MHz > 40 dB
20 MHz to 100 MHz > 30 dB
Noise Peak-to-Peak 3Q levels or 1 mV, whichever is greater
Common Mode Rejection Ratio 20 dB @ 50 MHz
XY Bandwidth
Bandwidth
Phase error @ 1 MHz
100 MHz
1.8 degrees
Performance Characteristics
HP 54645A/D Oscilloscopes
9-3

Table of Contents