EasyManua.ls Logo

HP 54645A - HP 54645 A;D Oscilloscope Trigger System

HP 54645A
301 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HP 54645A/D Oscilloscope Trigger System
Sources Channels 1 and 2, line, and EXT (on HP 54645A only)
Range (Internal) ± 6 div
Edges Either rising or falling
Sensitivity
< 10 mV/div, DC to 25 MHz
< 10 mV/div, 25 MHz to 100 MHz
> 10 mV/div, DC to 25 MHz
> 10 mV/div, 25 MHz to 100 MHz
1 div or 2 mV
1.5 div or 3 mV
.35 div or 3.5 mV
1 div or 10 mV
Coupling AC, DC, HF reject (~50 kHz), LF reject (~50 kHz), noise
reject
Maximum Input 400 V (DC + peak AC)
Range (External)
Sensitivity
Coupling
± 18 V
dc to 25 MHz, < 50 mv
25 MHz to 100 MHz, < 100 mV
DC, HF reject (~50 kHz), noise reject
Modes Auto, Auto level, Triggered (normal), Basic TV (Line,
Field), pattern/logic (on HP 54645D only).
Holdoff Time ~200 ns to ~25 seconds
Performance Characteristics
HP 54645A/D Oscilloscope Trigger System
9-7

Table of Contents