EasyManua.ls Logo

HP 54645A - Page 223

HP 54645A
301 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1
Check the rear panel
CALIBRATOR
output level.
If you are not sure how to check the
CALIBRATOR
, see "To check the output
of the
CALIBRATOR."
2
Load the default calibration factors.
a
Set the rear-panel
CALIBRATION
switch to the
UNPROTECTED
position.
b
Press
Print/Utility
. Then press
Service Menu
and
Self Cal Menu
.
c
Press the
Load Defaults
softkey.
Vertical
self cal
3
After the message "Default scope calibration factors have been
loaded" is displayed on the lower side of the display, press
Vertical
.
4
Follow the instructions on the display, then press
Continue
.
5
After the message "Press Continue to return to calibration menu"
appears on the display, press the
Continue
softkey
Delay self cal
for HP 54645A
Only
6
Connect a pulse generator set to 100 kHz and 1 V
p-p
, with a rise time
of less than 10 ns, to channels 1 and 2. Place the feedthrough at the
BNC tee. Make sure you use the HP 10502A cables to ensure equal
cable lengths.
7
Press the
Delay
softkey, then follow the instructions on the display.
8
Set the rear-panel
CALIBRATION
switch to
PROTECTED
.
Delay self cal
for HP 54645D
Only
9
Connect channels A1 and A2 to the front-panel probe adjust signal
(5 V
p-p
, 1.2 kHz) using two 10:1 probes (HP 10074A). Connect the
ground leads on the 10:1 probes to the front-panel ground lug using
the banana ground plug (HP 1252-7310).
10
Connect digital channel D0 to the front-panel probe adjust signal
(5 V
p-p
, 1.2 kHz) using the 16-channel cable (HP 54620-61801).
Connect the ground lead of this cable to the front-panel ground lug.
11
Press the Delay softkey, then follow the instructions on the display.
12
Set the rear-panel
CALIBRATION
switch to
PROTECTED
.
Testing, Adjusting, and Troubleshooting
To perform the self-calibration
7-34