UPDATES
Page
1-2:
After paragraph
1-9
add the following:
Manufacturer's Declaration
NOTE
NOTE
This is to certify that this product meets the radio
Hiqmit wird bescheinigt, dass dieses GeratlSystem
frequency interference requirements of Directive
in Ubereinstimmung mit den Bestimmungen von
FTZ
1046/1984. The German Bundespost has been
Pdstverfiigung 1046184 funkentst6x-t ist.
notified that this equipment was put into circulation
and has been granted the right to check the product
Der Deutschen Bundespost wurde das Inverkehr-
type for compliance with these requirements.
bringen dieses Geraes/Systems angezeigt und die
Berechtigung
zur Uberpriifung der Serie
auf
Ein-
haltung der Bestimmungen eingeraumt.
Note: If test and measurement equipment is oper-
ated with unshielded cables and/or used for mea-
Zusatlinfomation fiir Mess- und ~est~erate:
surements on open set-ups, the user must insure that
under these operating conditions, the radio
fie-
Werden Mess- Testgerate mit ungeschimten
quency interference limits
are
met at the border of
ICabeln
und/oder
in
offenen
Messaufbauten
ver-
his premises.
wendet, so ist vom
Betreiber sicherzustellen, dass
.
die
Funk-Entstorbestimmungen
unter Betriebs-
Model
HP
83522A
bedingungen an seiner Grundstucksgrenze ein-
gehalten werden.
Page 1-3, Table 1-1:
Under
"Power Variation,
Internally Leveled," add the following specification:
"With Option 002:
+
0.35
dB (in
0
dB attenuation step)."
Delete all references to Stability
With
Time (in a
10
minute period after one hour warmup).
Page 1-4, Table 1-2:
Add
the
following typical characteristic:
Power Variation
Internally Leveled (Option
002):
+
0.5
dB (in
10
dB to 70 dB attenuation steps).
Add STABILITY WITH TIME (in a
10
minute period after one hour warmup at the same frequency setting):
<
+
100
kHz.
Page
4-4,
Table
4-2:
Under
"4-14. Output Amplitude,
Internally Leveled:
2.25
dB," add the following:
"Opt.
002:
+
0.35
dB."
Page
4-5,
Table
4-2,
Section
4-15:
Delete
all
references to Time (10 minutes): specifications.
Page 5-1, Paragraph
5-1
1:
Replace Paragraph
5-12
with the following
Adjustment procedures are given in the proper sequence to allow for interrelated adjustments. However,
perform Paragraph
5-22
before Paragraph
5-21,
all other adjustments should
be
done in order of appearance.
Adjustments having to do with the leveling loop (paragraph
5-20
through
5-24)
are interactive and should
be
performed as a group.
UPDATES