Operation
FREQ CAL PROCEDURE
1. Press 8350A INSTR PRESET
CW
5
0
MHz
2.
Press
83522
a
(50 MHz
MARKER FREQUENCY
switches on
3. Adjust FREQ
CAL
@
control until MKR lamp
is
on.
4.
Press 8350A
1
0
5.
Press
83522A
10 MHz
@
Marker.
6.
Fine adjust FREQ CAL @control (if needed) until MKR lamp is on.
7.
Press
83522A
@
Marker.
8.
A small adjustment of FREQ
CAL@ control may be needed for MKR lamp to
light
Figure
3-5.
Front Panel FREQ
CAL
Procedure
3-7