100 Chapter2
Adjustment Procedures
10. Frequency Response Adjustment
10.On the HP 8560E/EC spectrum analyzer, adjust the RF Gain DAC
value using the front panel knob or keypad until the marker reads
10 dBm ±0.10 dB. Each DAC count yields an approximate 0.01 dB
change.
11.On the spectrum analyzer, press
to proceed to the next frequency.
12.Repeat steps 7 through 10 for all low-band frequencies 10 MHz.
13.If the firmware revision is later than 920528, perform steps 13
through 17. Otherwise, skip to step 18.
14.Press
until 10 MHz is displayed in the active function block.
Record the RF gain DAC value at 10 MHz.
10 MHz RF gain DAC value ______________________
15.Add 67 to the 10 MHz RF gain DAC value and record as the 2 MHz
RF gain DAC value.
2 MHz RF gain DAC value _______________________
16.Add 62 to the 10 MHz RF gain DAC value and record as the 6 MHz
RF gain DAC value.
6 MHz RF gain DAC value ________________________
17.Press until 2 MHz is displayed in the active function block. Use the
DATA keys to enter the 2 MHz RF gain DAC value recorded in step
14.
18.Press
until 6 MHz is displayed in the active function block. Use the
DATA keys to enter the 6 MHz RF gain DAC value recorded in step
15.
19.Press
PREV MENU STORE DATA, then YES on the HP 8560E/EC.
20.Place the WR PROT/WR ENA jumper on the A2 Controller assembly
in the WR PROT position.