Chapter 2 101
Adjustment Procedures
11. Calibrator Amplitude Adjustment
11. Calibrator Amplitude Adjustment
Assembly Adjusted
A15 RF assembly
Related Performance Test
Calibrator Amplitude and Frequency Accuracy
Description
The CAL OUTPUT amplitude is adjusted for 10.00 dBm measured
directly at the front panel CAL OUTPUT connector.
Figure 2-16 Calibrator Amplitude Adjustment Setup
Equipment
Measuring receiver ........................................................HP8902A
Power sensor ..................................................................HP8482A
Adapters
Type N (f) to BNC (m) ...................................................1250-1477