Chapter 2 119
Adjustment Procedures
17. 10 MHz Reference Adjustment-OCXO (Non-Option 103)
Figure 2-24 10 MHz Reference Adjustment Setup and Adjustment Location
Equipment
Frequency counter ......................................................HP5334A/B
Frequency standard..................... HP 5061B Cesium Beam Standard
(or any 10 MHz frequency standard with accuracy <±1 X 10
10
)
Cable
BNC, 122 cm (2 required).............................................HP10503A
Procedure
NOTE Failure to allow a 24 hour minimum warmup time for OCXO frequency
and temperature stabilization may result in oscillator misadjustment.