EasyManua.ls Logo

HP 8560E - Figure 10-1 YTO Loop Test Setup

HP 8560E
665 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 10 511
Synthesizer Section
Synthesizer Troubleshooting Section
16.Tune the source to 1 kHz less than the fractional N frequency. The
voltage measured on the DVM should be approximately 12 Vdc.
17.Tune the source to a frequency 1 kHz greater than the fractional N
frequency. The voltage measured on the DVM should be
approximately 12 Vdc.
18.If the DVM reading does not change, the A14 frequency control
assembly is defective. Reconnect W32 to A14J501.
Figure 10-1 YTO Loop Test Setup
Check A15 RF assembly (steps 18-25)
19.Disconnect W34 from A15U100J1 and disconnect W32 from
A15J101.
20.Connect a frequency counter to A15J101. Connect a high-frequency
test cable from an HP 8340A/B synthesized sweeper to A15U100J1.
See Figure 10-2 on page 512.
21.Connect a BNC cable from the HP 8560E/EC 10 MHz REF IN/OUT
to the HP 8340A/B FREQUENCY STANDARD EXT input.
22.Set the HP 8340A/B to the following settings: