Chapter 2 125
Adjustment Procedures
18. Tracking Oscillator Adjustment (Option 002)
14.Adjust A10C3 TRK OSC CTR until the frequency counter reads 300
MHz ±500 Hz.
15.Repeat steps 7 through 14 at least once more until no further
adjustment of A10C3 TRK OSC CTR is necessary.
16.Press LINE to turn the spectrum analyzer off. Replace the dust cap
screw on A10. Disconnect all cables from A10.
17.Reinstall A10 in the HP 8560E/EC.
Table 2-13 Tracking Oscillator Range Centering
N F1 (MHz) F2 (MHz) F
center
(MHz)
TRACK ADJ DAC Settings
COARSE FINE
1
2
3
4
5