188 Chapter3
Assembly Replacement
Procedure 7. A6A1 High Voltage Assembly (8560E only)
Ensure that the A16 FADC assembly is properly seated in the right
end of the slot.
9. Secure the A17 assembly (and A16 assembly in Option 007) with the
two flathead screws removed in step 18 under "Removal." See Figure
3-19 on page 187 (2).
10.For Option 007 spectrum analyzers: Connect the two mounting posts
to the left side frame using the two screws removed in step 17 under
"Removal." See Figure 3-19 on page 187 (1).
11.Fold the A2, A3, A4, and A5 assemblies into the spectrum analyzer
and secure the spectrum analyzer cover assembly as described in
"Procedure 5. A2, A3, A4, and A5 Assemblies."