36 Chapter1
General Information
Recommended Test Equipment
Sensors
Power sensor Frequency range: 10 MHz to 12 GHz HP 8481A*
P,A,T,
Maximum SWR:
M,V
1.40 (10 to 30 MHz)
1.18 (30 to 50 MHz)
1.10 (50 MHz to 2 GHz)
1.18 (2 to 12.4 GHz)
Power sensor Frequency range: 250 MHz to 350 MHz HP 8481D
P,A
Power range: 100 nW to 10 µW
Maximum SWR: 1.15 (250 to 350 MHz)
Power sensor Frequency range: 100 kHz to 2.9 GHz HP 8482A*
P,A,T,
Maximum SWR:
M,V
1.1 (1 MHz to 2.0 GHz)
1.30 (2.0 GHz to 2.9 GHz)
Other
Equipment
Controller Required to run operation verification software HP 9816A,
V
and adjustment/diagnostic software (HP 8564E/ HP 9836A/C,
HP 8565E) HP 310, 320
HP 332, 360
Oscilloscope Bandwidth (3 dB): dc to 100 MHz HP 54501A*
P,A,T
Two channels
Minimum vertical deflection factor: ≤5 mV/div
Minimum timebase setting: <100 ns
Digitizing display with time cursors
Delta-t cursor accuracy in 500 ns/Div: <0.1 µs
Amplifier Frequency range: HP 11975A
P
HP 8560E/EC, 2.0 to 2.9 GHz
HP 8561E, 2.0 to 6.5 GHz
HP 8563E, 2.0 to 8.0 GHz
HP 8564E, 2.0 to 8.0 GHz
HP 8565E, 2.0 to 8.0 GHz
Minimum output power (leveled)
2.0 to 8.0 GHz: +16 dBm
* Part of microwave workstation
P = performance tests; A = adjustments; M = test & adjustment module; T = troubleshooting;
V = operation verification
Table 1-3 Recommended Test Equipment
Instrument Critical Specifications for Equipment Substitution Recommended
Model
Use