418 Chapter8
IF Section
Automatic IF Adjustment
The references against which the Automatic IF Adjustment routine
aligns are:
10 MHz reference (A15)
Linear Scale Fidelity, especially the 10 dB gain stage in A4 Linear
Amplifier block.
15 dB Reference Attenuator (A5)
Cal Oscillator output power (A4)