62 Chapter2
Adjustment Procedures
1. High Voltage Power Supply Adjustment (8560E only)
Voltage marked on A6A1 HV Module = 0pt.......................... Vdc
6. Adjust A6R410 HV ADJ for a voltage equal to the voltage recorded
in step 5.
7. Press
LINE to turn the spectrum analyzer off and disconnect the
power cord. Wait at least 30 seconds for the high voltage power
supply capacitors to discharge.
8. Disconnect the DVM test leads from A6TP401 and A6TP405.
Reinstall the power supply cover.