Chapter 2 79
Adjustment Procedures
4. IF Amplitude Adjustments
6. On the HP 8560E/EC, reconnect W29 to A5J3. Press PRESET and set
the controls as follows:
Center frequency .....................................................300MHz
Span ........................ .........................................................0Hz
Reference level ............. ............................................10dBm
Resolution BW .............. ............................................300kHz
7. Connect a BNC cable between the HP 8560E/EC CAL OUTPUT
and INPUT 50.
8. On the HP 8560E/EC, press
MKR, CAL, and REF LVL ADJ.
9. Use the knob or step keys to adjust the REF LEVEL CAL setting
until the MKR reads 10.00 dBm ±0.1 dB.
10.On the HP 8560E/EC, press
STORE REF LVL.