Chapter 2 93
Adjustment Procedures
8. First LO Distribution Amplifier Adjustment
DAC value for 1.45 GHz = ____________________________
14.Set the "Sense EXT" value by pressing EXT LO LEVEL.
15.Use the knob or keypad to enter the DAC value for 1.45 GHz from
the band 0 sense voltage adjustment above.
16.Save the adjustment values by pressing
PREV MENU, STORE
DATA, and YES.
17.Move the jumper on A2J12 from WR ENA back to the WR PROT
position.