3-52 ES User’s Guide
Making Measurements
Measuring S11 Reflection Port 1 using a One-Port Calibration
Connect the DUT
Figure 3-19 Equipment Setup for an S
11
Measurement of a Two-Port Device
NOTE If you connect the DUT between both analyzer ports, it is recommended
that you use a 10 dB pad on the output of the DUT to improve
measurement accuracy. If you connect a two-port DUT to only one
analyzer port, it is recommended that you use a high-quality load (such
as a calibration standard) on the output of the DUT.