ES User’s Guide 2-13
Getting Started
Performing the Operator's Check
Make a Broadband Power Measurement
1. Leave the cable connected to the analyzer as shown in Figure 2-6.
NOTE The quality of the cable will affect these measurements; make sure you
use a cable with the characteristics described in “Equipment List” on
page 2-10.
2. Press
.
3. Press (unless done in the previous
measurement).
4. Verify that the data trace is within ±2 dB of 0 dBm. See Figure 2-8 for
a typical HP 8714ES result. The HP 8712ES should look similar, but
end at 1300 MHz.
Figure 2-8 Verify Broadband Power Measurement
MEAS 1
More
Power
FREQ
Start
10
MHz
SCALE 1
Enter
POWER 0
dBm

Table of Contents