EasyManua.ls Logo

HP 8712ES - To Use Relative Limits

HP 8712ES
507 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-36 ES User’s Guide
Using Instrument Functions
Using Limit Testing
To Use Relative Limits
There may be times when you are interested in the shape of a
measurement trace, but not concerned with the absolute amplitude. For
example, Figure 4-16 on page 4-31 shows limit lines created for tuning a
filter to a particular shape. If the shape is more important than the
amplitude, you can make the limit lines relative to the peak point of the
trace using the reference tracking function.
In this example, you would press
. The limit lines are now relative to the peak point on the
measurement trace rather than set at an absolute amplitude.
For more information on reference tracking, see “Using Reference
Tracking” on page 4-42.
SCALE
Reference Tracking
Track Peak