5-4 ES User’s Guide
Optimizing Measurements
Increasing Sweep Speed
To Increase the Start Frequency
Since the analyzer sweeps frequencies below approximately 20 MHz at a
slower rate, you can increase the start frequency to speed up the sweep.
1. Press .
2. Enter the highest start frequency possible for your measurement.
To Set the Sweep Time to AUTO Mode
Auto sweep time mode (the factory preset instrument mode) maintains
the fastest sweep speed possible for any particular measurement
settings.
1. Press and look at the softkey
label. When AUTO is all capital letters, it indicates that the analyzer
is in auto sweep time mode. If MAN is all capital letters, the analyzer
is in manual sweep time mode.
2. If necessary, press to toggle the time mode
to AUTO.
To Widen the System Bandwidth
Press to widen the IF bandwidth. As the
bandwidth increases, the sweep speed gets faster and the trace noise
increases.
Table 5-1 Relationship between System Bandwidth and Sweep Speed
FREQ
SWEEP
AVG
System Bandwidth Sweep Speed
Wide
Medium Wide
Medium
Medium Narrow
Narrow
Fine
fastest
faster
fast
slow
slower
slowest