6-6 ES User’s Guide
Calibrating for Increased Measurement Accuracy
Measurement Calibration Overview
To perform a quality calibration:
Use the highest quality standards available and take care of them.
When performing a user two-port, user one-port, or an enhanced
response calibration, be sure to select the correct cal kit by pressing
.
Make consistent connections of the standards and DUTs (if possible,
use a torque wrench when using 3.5 mm connectors).
Minimize temperature variations.
Use high-quality cables (semi-rigid cables, for example).
Minimize movement of cables.