7-8 ES User’s Guide
Front/Rear Panel
Connectors
Table 7-1 HP-IB Interface Cables Available
As many as 14 HP-IB instruments can be connected to the analyzer (15
total instruments in the system). The cables can be interconnected in a
star pattern (one central instrument with the HP-IB cables emanating
from that instrument like spokes on a wheel), or in a linear pattern (like
boxcars on a train), or a combination of the two. See Figure 1-5, “HP-IB
Connection Configurations, on page 1-13. There are certain restrictions
that must be followed when interconnecting instruments:
Each instrument must have a unique HP-IB address, ranging from 0
to 30.
In a two-instrument system that uses just one HP-IB cable, the cable
length must not exceed 4 meters (13.2 ft).
When more than two instruments are connected on the bus, the cable
length to each instrument must not exceed two meters (6.6 ft).
The total cable length between all instruments must not exceed
20 meters (66 ft).
Hewlett-Packard manufactures HP-IB extender instruments (Models
HP 37204A and HP 37204B) that overcome the range limitations
imposed by the cabling rules. These extenders allow twin pair cable
operation up to 1 km (3,280 ft), and telephone modem operation over any
distance. HP Sales and Service Offices can provide additional
information on the HP-IB extenders.
HP-IB Cable
Part Number
Length
HP 10833A 1 m (3.3 ft)
HP 10833B 2 m (6.6 ft)
HP 10833C 4 m (13.2 ft)
HP 10833D 0.5 m (1.6 ft)